Light incubator troubleshooting
Light incubator precautions and maintenance 1. After the climate box has landed, it should be leveled if the ground is uneven. Pet Toys YIWU CHUANGCHENG IMPORT&EXPORT CO.,LTD , https://www.tradingsix.com
2. During the handling of the climate chamber, it is forbidden to invert or tilt at a height greater than 45 degrees (since the unit is equipped with a compressor).
3. When there is no need for lighting in the box, the lighting switch should be placed in the "off" position to avoid affecting the upper temperature and prolong the life of the lamp.
4. Each time the water tank of the humidifier leaves the base, the water in the base must be inverted to prevent the water from overflowing.
5. When the climate tank is in normal operation, the contents of the tank should not be over-squeezed, and the air circulation inside the box should not be affected to ensure uniform temperature inside the box.
6. When the climate tank is not in use, it should be treated with moisture. The specific method is as follows: first clean the box, then set the temperature at 40 °C, run for 5 hours, and open the door every two hours to release the moisture. After the treatment is finished, Unplug the power cord and store it.
7. The tank liner and the surface of the tank should be wiped frequently to keep it clean and increase the transparency of the glass. Do not wipe the outer surface with acid, alkali or other corrosive solutions.
6. When using the humidifier, distilled water or purified water should be used. When there is sediment at the ultrasonic transducer, a few drops of washing essence should be added. After washing with a soft brush, the water is washed away. (It must be cleaned regularly for about 10 days)
8. If the climate box is not used for a long time, unplug the power cord. It should be operated regularly (2 to 3 days) on a regular basis (usually once in the first quarter) to remove moisture from the electrical parts and avoid damage to the equipment.
9. The maintenance of the climate box should be carried out by qualified personnel, and should be with our after-sales service center before maintenance.
10. If you need or ship it back for repair, please also contact our after-sales service center. When shipped back:
11. Please clean the climate box first to avoid injury to the maintenance personnel; attach a brief description of the fault error;
12. Keep the climate box vertical and upright during transportation;